著者
松田 忠義 吉田 翼 有岡 孝則 高塚 貞雄 二川 佳央 菊地 眞
出版者
Japanese Society for Thermal Medicine
雑誌
日本ハイパーサーミア学会誌 (ISSN:09112529)
巻号頁・発行日
vol.4, no.4, pp.317-329, 1988-12-01 (Released:2009-10-21)
参考文献数
21
被引用文献数
5 5

Heating experiments on phantoms and miniature pigs were made by using two lens applicators and a waveguide applicator fed by the 430MHz microwave heating system (HTS-100, producted by Tokyo Keiki Co., Ltd.). Temperature distributions of agar phantoms and miniature pigs were measured at regular intervals during heating. As the results of these experiments, the followings are obtained.1. The results of effective heating depth varying with a focal length confirm lens effect.2. Using a four-aperture lens applicator and miniature pigs, the heating depth of 6cm and it almost doubles that of the ordinary waveguide applicators. The heating area is 12cm x 6cm at 4cm depth.3. Dielectric properties of miniature pig buttocks are nearly equal to these of the agar phantom containing 0.35% NaCl and 0.05% NaN3.

言及状況

外部データベース (DOI)

Twitter (1 users, 1 posts, 1 favorites)

@7L3KPX どうも430MHzを使用した医療機器(加温装置)に関する日本語の論文が、1988年に出てるんですよね…何故430MHzなのかまでは分からないのですが… https://t.co/B00DmxwYqC