著者
畑川 政勝 井上 誠 庄垣 雅史 岸本 健治 吉田 梨影
出版者
公益社団法人日本放射線技術学会
雑誌
日本放射線技術學會雜誌 (ISSN:03694305)
巻号頁・発行日
vol.48, no.11, pp.1932-1938, 1992-11-01
被引用文献数
13

The modulation transfer function (MTF) has been used for evaluation of radiographic image quality. However, in digital radiographic system such as a computed radiography (CR), the MTF can not be used in the same way of an analog system for evaluation because digital system is not shift-invaliant system. Then, we suggest the "MTF_<final>" which includes geometric unsharpness, detector unsharpness, unsharpness of the sampling aperture and unsharpness of the display (Laser-printer). It is an analog component in the digital system and in agreement with the definition of the MTF. Furthermore, an overall aliasing has to be used with MTF_<final> for evaluating the digital system because it is one of the important factors of the digital system. The MTF_<final> and the overall aliasing can be obtained from the overall MTFs of the center alignment and the shifted alignment.

言及状況

Twitter (1 users, 4 posts, 0 favorites)

こんな論文どうですか? ディジタル系における MTF の検討 : ファイナル MTF およびオーバーオールエリアシング(畑川 政勝ほか),1992 http://t.co/81t1T4lGlz
こんな論文どうですか? ディジタル系における MTF の検討 : ファイナル MTF およびオーバーオールエリアシング(畑川 政勝ほか),1992 http://t.co/LQZvlsGA
こんな論文どうですか? ディジタル系における MTF の検討 : ファイナル MTF およびオーバーオールエリアシング,1992 http://ci.nii.ac.jp/naid/110003468089
こんな論文どうですか? ディジタル系における MTF の検討 : ファイナル MTF およびオーバーオールエリアシング,1992 http://ci.nii.ac.jp/naid/110003468089

収集済み URL リスト