著者
草場 志帆里 李 知炯 甲斐 智希 福本 悠斗 山越 健弘
出版者
福岡工業大学総合研究機構
雑誌
福岡工業大学総合研究機構研究所所報
巻号頁・発行日
no.1, pp.97-102, 2018-12

The aim of this study is to compare the heart rate (HR) derived from electrocardiogram (ECG) with the pulse rate (PR) derivedfrom the green light ear photo-plethysmogram (PPG) using an earphone. We developed a compact device that the PPG sensorcomprising a green LED and phototransistor was embedded in the earphone. The weight of device was 8.8 g. In 10 subjects,simultaneous measurements of ECG, and green light (525 nm) PPG from the front of tragus were made at the following treadmillspeeds: 0, 2, 4, 6, 8 km/h. During walking (2, 4 km/h), the results showed that the PR derived from the green light PPG from theear significantly related to HR derived from ECG (r = 0.98). In addition, the limit of agreement in Bland-Altman plots betweenthese indices were ± 3.95 (1.96 SD) bpm. In conclusion, these findings suggest that the green light ear PPG using earphone mightbe practical measurement for monitoring HR during walking.

言及状況

Twitter (1 users, 2 posts, 0 favorites)

こんな論文どうですか? スマートイヤーモニターシステムの開発研究-耳部脈拍数検出のためのイヤホン型高感度緑光電容積脈波計測システム開発-(草場 志帆里ほか),2018 https://t.co/iG1V6Papm8 The aim of this …
こんな論文どうですか? スマートイヤーモニターシステムの開発研究-耳部脈拍数検出のためのイヤホン型高感度緑光電容積脈波計測システム開発-(草場 志帆里ほか),2018 https://t.co/iG1V6PsyAg The aim of this …