著者
中川 靖夫 大久保 和明 堀内 元美 戸沢 均 小林 健二 堀江 和泉 千田 直道
出版者
社団法人照明学会
雑誌
照明学会誌 (ISSN:00192341)
巻号頁・発行日
vol.82, no.8A, pp.588-593, 1998

An intercomparison of relative spectral responsivity measurements of illuminance meters was carried out among five companies-TOPCON, HIOKI, Minolta Co., Ltd., Yokogawa M&C Corp.and Matushita Electric Industrial Co., Ltd. The spectral responsivity of the illuminance meters were measured by each company's spectral responsivity measurement system calibrated to each company's standard. As a result, the measured values of f_s(f_s is error definition of relative spectral responsivity recommended by Japanese Industrial Standard JIS-C-1609)were different within 1.7. The f_s is identical with f_1' recommended by CIE.