- 著者
 
          - 
             
             水島 茂喜
             
             植木 正明
             
             根津 嘉明
             
             大岩 彰
             
          
 
          
          
          - 出版者
 
          - The Vacuum Society of Japan
 
          
          
          - 雑誌
 
          - 真空 (ISSN:05598516)
 
          
          
          - 巻号頁・発行日
 
          - vol.43, no.3, pp.181-183, 2000-03-20 (Released:2009-10-20)
 
          
          
          - 参考文献数
 
          - 10
 
          
          
        
        
        
        A new electronic mass comparator, capable of automatic weighing in vacuum, was introduced into the National Research Laboratory of Metrology. The standard deviation of 0.1 μg, a factor of 3 smaller than the standard deviation of the best previous measurement using a mechanical balance, was achieved for the mass comparison between two 1-kg weights. For eight days, mass difference of two 1-kg weights with the surface area difference of 66.7 cm2 was measured in vacuum. We observed an outgassing rate of 9×10-6 Pa·3·s-1·m-2 after an evacuation for 3.9-6.8 hours, and below 6×10-9 Pa·m3·s-1·m-2 after an evacuation for 107 hours. Residual gas in the vacuum chamber, which housed the electronic mass comparator, was analyzed to monitor the oil contamination. At a total pressure of 5.5×10-4 Pa, we observed the main peaks due to hydrocarbon at 41, 43, and 57 amu with a partial pressure of 3×10-6 Pa.