著者
Yoshikawa M. Yasuhara R. Morimoto M. Shima Y. Kohagura J. Sakamoto M. Nakashima Y. Imai T. Yamada I. Kawahata K. Funaba H. Minami T.
出版者
American Institute of Physics
雑誌
Review of scientific instruments (ISSN:00346748)
巻号頁・発行日
vol.83, no.10, pp.10E333, 2012-10
被引用文献数
17 6

In the GAMMA 10 tandem mirror, the typical electron density is comparable to that of the peripheral plasma of torus-type fusion devices. Therefore, an effective method to increase Thomson scattering (TS) signals is required in order to improve signal quality. In GAMMA 10, the yttrium-aluminum-garnet (YAG)-TS system comprises a laser, incident optics, light collection optics, signal detection electronics, and a data recording system. We have been developing a multi-pass TS method for a polarization-based system based on the GAMMA 10 YAG TS. To evaluate the effectiveness of the polarization-based configuration, the multi-pass system was installed in the GAMMA 10 YAG-TS system, which is capable of double-pass scattering. We carried out a Rayleigh scattering experiment and applied this double-pass scattering system to the GAMMA 10 plasma. The integrated scattering signal was made about twice as large by the double-pass system.