著者
Masao Kato Shoichi Asakura Hideaki Kimoto Tatsuhiko Sasaki Ko Dezawa Toshihiko Amemiya Kunihito Matsumoto Yoshinori Arai
出版者
Nihon University School of Dentistry
雑誌
Journal of Oral Science (ISSN:13434934)
巻号頁・発行日
pp.23-0298, (Released:2023-11-30)
参考文献数
24
被引用文献数
1

Purpose: To evaluate the image quality of vertical dual-exposure panoramic radiography (PR), which merges two PR images taken at different focus heights to reduce ghost images of cervical vertebrae (CV) and intervertebral spaces (IVS) in the incisor region.Methods: PR images of an aluminum block, a CV phantom and a human head phantom were taken at 0 mm and merged with and subtracted from PR images taken at other heights (0, 5, 10, 15, and 20 mm) to create new images, e.g., Merg0 + 15 mm and Sub0 − 10 mm. The subtracted images were analyzed subjectively according to the uniformity on the line profile. Merged images were evaluated subjectively by six raters to determine the influence of the ghost images.Results: Objective evaluation revealed a positional shift in the ghost images according to the height of the focus for both phantoms. In the subjective evaluation, the normal PR (Merg0 + 0 mm) showed the worst score, indicating strong influence of CV and IVS ghost images.Conclusion: The vertical dual-exposure PR method, which merges PR images taken at the normal position and a higher X-ray focus, can reduce CV and IVS ghost images in the incisor region.

言及状況

外部データベース (DOI)

Twitter (1 users, 1 posts, 0 favorites)

【論文】Masao Kato. Reduction of cervical vertebra ghost images in panoramic radiography using vertical dual exposure. https://t.co/kZ8THoqqsr