著者
藤村 一郎 市川 勝弘 寺川 彰一 原 孝則 三浦 洋平
出版者
公益社団法人 日本放射線技術学会
雑誌
日本放射線技術学会雑誌 (ISSN:03694305)
巻号頁・発行日
vol.67, no.6, pp.640-647, 2011-06-25 (Released:2011-07-01)
参考文献数
20

For head computed tomography (CT), non-helical scanning has been recommended even in the widely used multi-slice CT (MSCT). Also, an acute stroke imaging standardization group has recommended the non-helical mode in Japan. However, no detailed comparison has been reported for current MSCT with more than 16 slices. In this study, we compared the non-helical and helical modes for head CT, focusing on temporal resolution and motion artifacts. The temporal resolution was evaluated by using temporal sensitivity profiles (TSPs) measured using a temporal impulse method. In both modes, the TSPs and temporal modulation transfer factors (MTFs) were measured for various pitch factors using 64-slice CT (Aquilion 64, Toshiba). Two motion phantoms were scanned to evaluate motion artifacts, and then quantitative analyses for motion artifacts and helical artifacts were performed by measuring multiple regions of interest (ROIs) in the phantom images. In addition, the rates of artifact occurrence for retrospective clinical cases were compared. The temporal resolution increased as the pitch factor was increased. Remarkable streak artifacts appeared in the non-helical images of the motion phantom, in spite of the equivalent effective temporal resolution. In clinical analysis, results consistent with the phantom studies were shown. These results indicated that the low pitch helical mode would be effective for emergency head CT with patient movement.
著者
井上 健 市川 勝弘 原 孝則 瓜倉 厚志 星野 貴志 三浦 洋平 寺川 彰一 宇都 文昭
出版者
公益社団法人 日本放射線技術学会
雑誌
日本放射線技術学会雑誌 (ISSN:03694305)
巻号頁・発行日
vol.68, no.12, pp.1631-1636, 2012-12-20 (Released:2012-12-20)
参考文献数
12
被引用文献数
3 1

Iterative reconstruction methods can reduce the noise of computed tomography (CT) images, which are expected to contribute to the reduction of patient dose CT examinations. The purpose of this study was to investigate impact of an iterative reconstruction method (iDose4, Philips Healthcare) on vessel visibility in coronary CT angiography (CTA) by using phantom studies. A simulated phantom was scanned by a CT system (iCT, Philips Healthcare), and the axial images were reconstructed by filtered back projection (FBP) and given a level of 1 to 7 (L1-L7) of the iterative reconstruction (IR). The vessel visibility was evaluated by a quantitative analysis using profiles across a 1.5-mm diameter simulated vessel as well as visual evaluation for multi planar reformation (MPR) images and volume rendering (VR) images in terms of the normalized-rank method with analysis of variance. The peak CT value of the profiles decreased with IR level and full width at half maximum of the profile also decreased with the IR level. For normalized-rank method, there was no statistical difference between FBP and L1 (20% dose reduction) for both MPR and VR images. The IR levels higher than L1 sacrificed the spatial resolution for the 1.5-mm simulated vessel, and their visual vessel visibilities were significantly inferior to that of the FBP.