著者
和田森 直 石原 康利
出版者
公益社団法人 日本生体医工学会
雑誌
生体医工学 (ISSN:1347443X)
巻号頁・発行日
vol.49, no.1, pp.220-225, 2011-02-10 (Released:2011-12-13)
参考文献数
31

A measurable depth of the blood glucose level by using photoacoustic spectroscopy (PAS) was discussed experimentally in order to improve detection sensitivity. Since a measurable depth of the PAS based on a modulation frequency of the chopped light irradiated to a sample, the relationship between modulation frequency and the thickness of a sample was evaluated. In our experiments, the photoacoustic detector consisted of an acoustic resonant pipe and an optical microphone, and a two-layer model composed of silicone sheets with the different optical absorption were used. We varied the thickness of the upper-layer for the two-layer model from 0.50 to 5.00 mm at each modulation frequencies to investigate the relationship between the measurable depth and the modulation frequency. As results, it is shown that the measurable depth was 2-3 mm with modulation frequency of 1.6-1.9 kHz. Furthermore, the reason for the measurable depth to be deeper than the thermal diffusion length of a sample was discussed. From these analyses of thermoelastic wave, the relation between the photoacoustic signal propagation and the measurable depth in a tissue was clarified.
著者
和田森 直
出版者
長岡技術科学大学
雑誌
挑戦的萌芽研究
巻号頁・発行日
2015-04-01

振動として可聴信号を耳周辺軟骨に加えるとその可聴信号を知覚できる現象を利用して、断続光の吸収に伴う発熱により、試料内部で振動が発生する光音響効果を応用した骨導音声情報提示装置を開発した。耳周辺軟骨に類似したゴム・シートから発生したPA信号の力のレベルは、日本工業規格に示されている骨導受話器を乳突部に装着した場合の基準等価いき値の力のレベルを最大47dB程度下回った。生体内の主な光吸収物質である水やヘモグロビンは、軟骨に比べおよそ100倍の吸光度を持つ。水やヘモグロビンの吸収波長付近の光源を断続光に用いることにより、基準いき値以上にPA信号の力のレベルを向上できる。