- 日本放射線技術学会雑誌 (ISSN:03694305)
- 巻号頁・発行日
- vol.72, no.7, pp.581-588, 2016 (Released:2016-07-20)
- 参考文献数
- 25
- 被引用文献数
-
1
This study was conducted to measure the recognition time of the test pattern and to investigate the effects of the maximum luminance in a medical-grade liquid-crystal display (LCD) on the recognition time. Landolt rings as signals of the test pattern were used with four random orientations, one on each of the eight gray-scale steps. Ten observers input the orientation of the gap on the Landolt rings using cursor keys on the keyboard. The recognition times were automatically measured from the display of the test pattern on the medical-grade LCD to the input of the orientation of the gap in the Landolt rings. The maximum luminance in this study was set to one of four values (100, 170, 250, and 400 cd/m2), for which the corresponding recognition times were measured. As a result, the average recognition times for each observer with maximum luminances of 100, 170, 250, and 400 cd/m2 were found to be 3.96 to 7.12 s, 3.72 to 6.35 s, 3.53 to 5.97 s, and 3.37 to 5.98 s, respectively. The results indicate that the observer’s recognition time is directly proportional to the luminance of the medical-grade LCD. Therefore, it is evident that the maximum luminance of the medical-grade LCD affects the test pattern recognition time.