著者
川崎 達哉 加藤 英幸 笠原 哲治 田岡 淳一 梅北 英夫 桝田 喜正
出版者
公益社団法人 日本放射線技術学会
雑誌
日本放射線技術学会雑誌 (ISSN:03694305)
巻号頁・発行日
vol.76, no.3, pp.278-284, 2020 (Released:2020-03-20)
参考文献数
18

This study was designed to clarify the relation between the pressure resistance of an angiographical tube and the amount of contrast medium injected under a connected microcatheter used for interventional radiology (IVR). We investigated the injection pressure and the expansion rate at the center of the tube during contrast enhancement by setting the power injector to 1200 PSI pressure, with 2.0 ml/s injection speed, 10 ml injection volume, 5.0 s injection time, and 0 s rise time for tubes with different pressure resistance performance (low or high). Then we examined the amount of contrast medium material discharged from the microcatheter. The low-pressure resistant tube (less than 140 PSI) injection pressure exceeded the pressure performance. The expansion rate increased to 49%, presenting a risk of rupture. The injection pressure of the high-pressure resistant tube (less than 1200 PSI) was within the pressure-resistance performance. The expansion rate increased to 38%. However, when the contrast medium discharge amount contributing to the image was measured within the injection time under the condition of 10 ml injection for 5.0 s, the former was 2.3 ml and the latter was 4.2 ml. The entire amount was not discharged during the injection period. It became apparent that it is discharged in drips after some time. Results show that the tube expansion caused retention of the contrast medium inside, which decreases the actual amount of the injected contrast medium. From the results, we infer the possibility of preventing reduction of the injected contrast medium amount attributable to expansion.