著者
中村 忍 川田 秀道 黒木 英郁 溝口 明日実
出版者
公益社団法人 日本放射線技術学会
雑誌
日本放射線技術学会雑誌 (ISSN:03694305)
巻号頁・発行日
vol.71, no.11, pp.1096-1102, 2015 (Released:2015-11-20)
参考文献数
21
被引用文献数
2

We evaluated the effect of orthopedic-metal artifact reduction (O-MAR) for metal artifact in computed tomography with 73 simulated seeds for brachytherapy in different sizes of display field of view (DFOV) obtained by helical scan under the same clinical scan condition. The metal artifacts were analyzed with the Gumbel's method by changing DFOV sizes 80 mm, 160 mm, and 320 mm. Gumbel distribution, scale parameter (γ), and location parameter (β) of the metal artifacts with O-MARwere compared with that of the metal artifacts with filtered back projection (FBP). In conclusion, it was considered that the effect of metal artifact reduction with O-MARwas influenced by DFOV size in this study. The reduction rates of scale parameter (γ) were 22.3%, 21.3%, and 10.0% in DFOV 80 mm, 160 mm, and 320 mm, respectively. The reduction rates of location parameter (β) were 27.4%, 23.4 %, and 9.8%. Therefore, the effect of metal artifact reduction with O-MARshowed the tendency of increasing with decreasing DFOV size.