- 著者
-
Bui Kamihoriuchi
Yoichi Otsuka
Aya Takeuchi
Futoshi Iwata
Takuya Matsumoto
- 出版者
- The Mass Spectrometry Society of Japan
- 雑誌
- Mass Spectrometry (ISSN:2187137X)
- 巻号頁・発行日
- vol.7, no.2, pp.S0078, 2019-03-06 (Released:2019-03-07)
- 参考文献数
- 22
- 被引用文献数
-
1
7
Ambient sampling and ionization techniques based on direct liquid extraction and electrospray ionization are of great value for rapid analysis and mass spectrometry imaging. Scanning probe electrospray ionization (SPESI) enables the sampling and ionization of analyte molecules in a solid material using a liquid bridge and electrospray, respectively, from a single capillary probe. To further improve SPESI, it is essential to understand the dynamic behavior of nanoliter volumes of liquids during sampling and ionization. In this study, the dynamic formation and breakage of the liquid bridge and the subsequent electrospray ionization were investigated by measuring the displacement of the capillary probe using a new optical technique. Measurements revealed that both the time from the formation of the liquid bridge to its breakage and the time from the breakage of the liquid bridge to the detection of analyte ions were correlated with the physical properties of the solvent. It was also found that both of these times were positively correlated with the flow rate. These results will not only lead to the improvement of sampling and ionization efficiencies but also afford a greater understanding of the physicochemical properties of charged nanoliter volumes of liquids.