著者
Ippei SUZUKI Yasuhito KUMAI Machiko KITAGAWA Yuka KISHIMOTO Keizo UMEGAKI Tsuyoshi CHIBA Jun TAKEBAYASHI
出版者
The Japan Society for Analytical Chemistry
雑誌
Analytical Sciences (ISSN:09106340)
巻号頁・発行日
vol.35, no.11, pp.1269-1274, 2019-11-10 (Released:2019-11-10)
参考文献数
11
被引用文献数
3

In HPLC analyses of soluble dietary fiber, desalting processes using open, mixed-bed ion-exchange columns are time-consuming and labor-intensive. We developed and validated a simple desalting method using tandem cation/anion exchange SPE cartridges. We found that combining Bond Elut Jr SCX (upstream) and Bond Elut PSA (downstream) cartridges provided adequate desalting of test solutions. The developed method was then validated in an inter-laboratory study. Five test samples were prepared by mixing food matrixes with purified soluble dietary fiber and treated to generate solutions to test the desalting process. These solutions were then analyzed by eight different laboratories. The results demonstrated that the developed method is simple and reliable for desalting samples containing 140 to 945 mg/100 mL of soluble dietary fiber in preparation for HPLC analysis of soluble dietary fiber.