著者
Rinaldo Roberto de Jesus Guirro Elaine Caldeira de Oliveira Guirro Natanael Teixeira Alves de Sousa
出版者
The Society of Physical Therapy Science
雑誌
Journal of Physical Therapy Science (ISSN:09155287)
巻号頁・発行日
vol.26, no.4, pp.557-562, 2014 (Released:2014-04-23)
参考文献数
17
被引用文献数
3 7

[Purpose] Although shortwave diathermy has been widely used by physiotherapists, there are a few studies assessing the performance of the equipment in use. The aim of the present study was to evaluate the procedures adopted by physiotherapists as users of shortwave diathermy continuous (CSWD), as well as to measure the power output and frequency of CSWD equipment. [Subjects and Methods] Twenty-three physical therapists were interviewed and 23 CSWD equipment were evaluated. Admeasurement was carried out by using a standard phantom to simulate the electrode-skin distance, which ranged from 0.5 to 3.0 cm. Data analysis was performed by using descriptive statistics, ANOVA, and a post-hoc Tukey’s test or Pearson’s correlation coefficient. [Results] The questionnaires showed that 48% of the interviewees use the correct electrode-skin distance, 70% use a single electrical outlet, and 35% use a grounded electrical outlet, and that 48% of the physiotherapy tables and 61% of the plinths were made of wood. However, only 13% of the interviewees perform yearly preventive maintenance. The highest power (95.56 W) was achieved at electrode-skin distances ranging from 1.0 to 1.5 cm, with distances of 2.5 cm and 3.0 cm being null in four and eight equipment, respectively. There was a negative correlation between power output and electrode-skin distance as well as between power output and purchase date. [Conclusion] The physiotherapists involved in this study had inadequate knowledge about the correct use of CSWD equipment, which may adversely affect its performance and patient safety.