- 著者
-
Isao Shitanda
Tasuku Yamaguchi
Yoshinao Hoshi
Masayuki Itagaki
- 出版者
- (社)日本化学会
- 雑誌
- Chemistry Letters (ISSN:03667022)
- 巻号頁・発行日
- pp.130681, (Released:2013-08-24)
- 被引用文献数
-
17
A fully screen-printed electrode chip was fabricated for glucose detection. The utility of the screen-printed electrode chip based on a Japanese paper was demonstrated through the determination of glucose using the reaction of glucose oxidase with tetrathiafulvalene. The calibration curve showed good linearity in the concentration range 10 to 100 mM, with a slope of 0.055 μA mM−1. The paper-based electrode chip is potentially applicable for point-of-care monitoring because it is easy to use, inexpensive, and readily disposable.