著者
Yoshimitsu Soga Osamu Iida Masahiko Fujihara Daizo Kawasaki Shigeru Saito Kazushi Urasawa Hiroyoshi Yokoi Eric J. Fernandez Jia Guo Masato Nakamura on behalf of the IN.PACT Japan Post-Market Surveillance Study
出版者
The Japanese Circulation Society
雑誌
Circulation Journal (ISSN:13469843)
巻号頁・発行日
pp.CJ-21-0491, (Released:2021-10-06)
参考文献数
27
被引用文献数
4

Background:To confirm the safety and efficacy of the IN.PACT Admiral drug-coated balloon (DCB) based on the indication approved by the Pharmaceuticals and Medical Devices Agency Japan in real-world patients with femoropopliteal artery disease.Methods and Results:IN.PACT PMS Japan was a prospective, multicenter, single-arm, post-market surveillance (PMS) study conducted in Japan that enrolled 304 participants (mean age 75.3±7.9 years). The primary endpoint was primary patency at 6 months following the index procedure, defined as freedom from clinically driven target lesion revascularization (CD-TLR) and freedom from restenosis as determined by duplex ultrasound (DUS) peak systolic velocity ratio (PSVR) ≤2.4 (assessed by the independent DUS core laboratory). Secondary endpoints included acute outcomes, primary patency at 12 months post-index procedure, freedom from CD-TLR, and major adverse events at 12 months. The mean lesion length was 97.81±58.97 mm. The primary endpoint, 6-month primary patency, was 91.3% (240/263). Kaplan-Meier estimates of primary patency and freedom from CD-TLR through 12 months were 91.5% and 94.1%, respectively. The CD-TLR rate was 5.8% (14/240) with low rates of thrombosis (0.8%) and target limb amputation (0.4%) at 12 months.Conclusions:The results of this real-world PMS study were consistent with outcomes from previous IN.PACT DCB studies, confirming the safety and efficacy of the IN.PACT Admiral DCB for broader use in patients seen in everyday practice.