著者
久野 弘明 高島 充 大川井 宏明
出版者
一般社団法人 電気学会
雑誌
電気学会論文誌C(電子・情報・システム部門誌) (ISSN:03854221)
巻号頁・発行日
vol.124, no.4, pp.935-940, 2004 (Released:2004-07-01)
参考文献数
12
被引用文献数
1 3

In this study, the possibility of the measurement of respiration, heart beat, and body movement on a bed was examined using the dynamic air-pressure sensor aiming at a daily health monitoring. The dynamic air-pressure sensor measures vital information using a change of air pressure. Twelve healthy volunteers participated in this study. The dynamic air-pressure sensor was installed under the bed mat and respiration and heart beat information were measured. This information was compared with the standard waveforms obtained from respiratory belt transducer and the electrocardiograph. As a result, both waveforms demonstrate a high correlation, and confirmed the validity of this method. A change of waveform and a quantitative evaluation of respiration, heart beat, and body movement measured from during sleep using this sensor can be useful for a daily health monitoring.