- 著者
-
Ryosuke Tomio
Tsuyoshi Uesugi
Kazunori Akaji
- 出版者
- 特定非営利活動法人 日本脳神経血管内治療学会
- 雑誌
- Journal of Neuroendovascular Therapy (ISSN:18824072)
- 巻号頁・発行日
- vol.14, no.3, pp.85-89, 2020 (Released:2020-03-20)
- 参考文献数
- 4
- 被引用文献数
-
1
Objective: The optimal heating temperature and time for the Echelon10 and Excelsior SL-10 microcatheters using a heat gun was investigated. The durability of the microcatheters after heat gun shaping for the second and third times was also examined.Methods: HAKKO FV-310 was used as the heat gun in this study. This heat gun can be set to 115°C, 125°C, and others. We measured the temperature at 2.5 cm from the nozzle of the heat gun. The Echelon10 and SL-10 microcatheters were shaped under two temperature conditions (115°C and 125°C) and three heating times (30 sec, 60 sec, and 90 sec). The microcatheter shape before heating had twice the curvature of the targeted shape.Results: The temperatures at 2.5 cm from the nozzle were 120.6°C and 127.8°C with the heat gun set at 115°C and 125°C, respectively. There was no macroscopic difference in the results of heat gun shaping of the Echelon10 among temperature settings (115°C and 125°C) or heating times (30 sec, 60 sec, and 90 sec). As degeneration of the heated tip of the SL-10 at 125°C occurred in four of five trials, heat gun shaping was performed using the 115°C setting. There was no macroscopic difference in the results of heat gun shaping of the SL-10 among heating times. Shaping for the second and third times was successful at 115°C and 30-sec heating time.Conclusions: The Echelon10 and SL-10 can be successfully shaped from twice the curvature of the targeted shape using a heat gun at 120°C for 30 sec. Shaping for the second and third times was successful using the same settings. Degeneration of the SL-10 was noted at temperatures above 130°C.