著者
Marimo MATSUMOTO Masaya TAMURA
出版者
The Institute of Electronics, Information and Communication Engineers
雑誌
IEICE TRANSACTIONS on Electronics (ISSN:09168516)
巻号頁・発行日
vol.E105-C, no.9, pp.398-406, 2022-09-01
被引用文献数
2

Couplers in a film-type capacitive wireless power charging (CWC) system for an implantable medical device were designed and analyzed in this work. Due to the high conductivity of the human body, two paths contribute to the power transmission, namely a high-frequency current and an electric field. This was confirmed by an equivalent circuit of the system. During analysis of the system, we used pig skin with subcutaneous fat, which has a high affinity with the human body, to search for a highly efficient electrode shape. Subsequently, we fabricated the designed coupler and measured ηmax. An ηmax of 56.6% was obtained for a half-circular coupler with a radius of 20 mm and a distance of 10 mm between adjacent couplers. This study will contribute to the realization of implantable devices that can be recharged during breaks or while sleeping at home and is expected to significantly reduce the burden on patients.