- 著者
-
田中 耕一
吉野 健一
内藤 康秀
豊田 岐聡
- 出版者
- 一般社団法人 日本質量分析学会
- 雑誌
- Journal of the Mass Spectrometry Society of Japan (ISSN:13408097)
- 巻号頁・発行日
- vol.56, no.2, pp.49-54, 2008-04-01 (Released:2008-04-15)
- 被引用文献数
-
2
2
Linear mode, reflectron mode, and multi-turn are techniques for m/z analysis in time-of-flight mass spectrometry. Reflectron mode and multi-turn enable analysis to be performed with a higher mass resolving power than that attained with linear mode. In this commentary, starting from the basic principles of time-of-flight mass spectrometry in linear mode, the possible improvements in mass resolving power that can be made by reflectron mode and multi-turn techniques will be explained.