著者
田中 耕一 吉野 健一 内藤 康秀 豊田 岐聡
出版者
一般社団法人 日本質量分析学会
雑誌
Journal of the Mass Spectrometry Society of Japan (ISSN:13408097)
巻号頁・発行日
vol.56, no.2, pp.49-54, 2008-04-01 (Released:2008-04-15)
被引用文献数
2 2

Linear mode, reflectron mode, and multi-turn are techniques for m/z analysis in time-of-flight mass spectrometry. Reflectron mode and multi-turn enable analysis to be performed with a higher mass resolving power than that attained with linear mode. In this commentary, starting from the basic principles of time-of-flight mass spectrometry in linear mode, the possible improvements in mass resolving power that can be made by reflectron mode and multi-turn techniques will be explained.