著者
鈴木 新 劉 和輝 金井 伸行
出版者
一般社団法人 電気学会
雑誌
電気学会論文誌C(電子・情報・システム部門誌) (ISSN:03854221)
巻号頁・発行日
vol.131, no.9, pp.1559-1563, 2011-09-01 (Released:2011-09-01)
参考文献数
16

This paper proposes a method for estimation of a body temperature by using a relation between the upper arm temperature and the atmospheric temperature. Conventional method has measured by armpit or oral, because the body temperature from the body surface is influenced by the atmospheric temperature. However, there is a correlation between the body surface temperature and the atmospheric temperature. By using this correlation, the body temperature can estimated from the body surface temperature. Proposed method enables to measure body temperature by the temperature sensor that is embedded in the blood pressure monitor cuff. Therefore, simultaneous measurement of blood pressure and body temperature can be realized. The effectiveness of the proposed method is verified through the actual body temperature experiment. The proposed method might contribute to reduce the medical staff's workloads in the home medical care, and more.