- 著者
-
大田 明博
臼田 孝
中野 英俊
- 出版者
- 一般社団法人 電気学会
- 雑誌
- 電気学会論文誌E(センサ・マイクロマシン部門誌) (ISSN:13418939)
- 巻号頁・発行日
- vol.123, no.6, pp.203-208, 2003 (Released:2003-09-01)
- 参考文献数
- 8
The calibration procedures of charge amplifier (CA) on vibration acceleration standard are described. The calibration principle and fundamental calibration procedures are explained in detail. Typical commercial CAs (3 types, 4 CAs) are evaluated in terms of the frequency sensitivity gain, the long-term stability, the linearity for input charge, the linearity for gain setting and the effects of environmental conditions (temperature and humidity). The maximum discrepancy in the sensitivity gain is approximately 4 % among the tested CAs. A relative sensitivity gain difference of 1.5 % still remains when CAs of the same type are compared. It should be mentioned that a difference of 1.5 % is not negligible. Such a large difference has not been commonly observed up to the present time. Based on these experimental results, the technical guides for charge amplifier calibration are summarized.