著者
喜屋 武典 大城 進 亀谷 浩昌 藤本 勝喜 赤崎 満 古謝 景春
出版者
日本医療薬学会
雑誌
病院薬学 (ISSN:03899098)
巻号頁・発行日
vol.19, no.1, pp.18-23, 1993-02-20

The use of a reagent for digoxin assay of fluorescence polarization immunoassay with TDX analyzer in the determination of serum deslanoside concentration was studied. Deslanoside calibrators with the concentration of 0.0-6.0ng/ml were prepared. The assay item for digitoxin was used for deslanoside after modification of the parameters. The coefficients of variation of within-run assay and between-run assay in this method were less than 9% at various concentrations tested. By use of a ultrafiltration method, the protein binding percent of deslanoside was determined to be zero or very small. In a patient with renal and hepatic failures, the half life (t1/2) and the apparent volume of distribution (Vd) were calculated to be 131 hours and 4.1l/kg after intravenous administration of deslanoside, respectively. On the other hand, an equation of Y=1.25X+0.08 was obtained from a correlation between digoxin concentration (X) given by normal digoxin assay and concentration of deslanoside (Y). Therefore, with a simple method, one can calculate the deslanoside concentrations.