- 著者
-
雨宮 正欣
長井 辰男
- 出版者
- 日本法科学技術学会
- 雑誌
- 日本鑑識科学技術学会誌 (ISSN:13428713)
- 巻号頁・発行日
- vol.4, no.2, pp.95-98, 2000 (Released:2010-02-05)
- 参考文献数
- 14
- 被引用文献数
-
1
3
Though easiness and availability of a screening test are important factors, its correctness is most important for a drug abuser's arrest according to screening results. The use of Triage® in screening tests to detect stimulants was examined. Abusers' urine samples contained methamphetamine at the concentration of which ranges from 1μg/ml to 479 μg/ml, and 66 μg/ml(n=68) on the average. Triage® showed negative reaction at 1000 μg/ml to 1500 μg/ml, even though there was a high concentration of methamphetamine. Triage® gives false negative result, even when there is a very high concentration of stimulant present in the urine. We should take this into consideration.