著者
Yasuki Uchida Yasuhiro Namura Mizuki Inaba Ayaka Osada Tasku Charleston-Coad Yoshiki Nakamura Mitsuru Motoyoshi
出版者
Nihon University School of Dentistry
雑誌
Journal of Oral Science (ISSN:13434934)
巻号頁・発行日
vol.63, no.3, pp.270-274, 2021 (Released:2021-06-29)
参考文献数
26
被引用文献数
5

Purpose: The aim of this study was to investigate the stability of orthodontic anchor screws (OASs) in the mid-palatal area according to pre-drilling diameter.Methods: The success rate of 161 OASs (83 patients, φ2.0 mm, 6.0 mm in length) placed in a corresponding area to the mesial and distal borders of the first molar (mesial zone and distal zone) was assessed according to placement location and pre-drilling diameter (1.2 and 1.5 mm). Placement torque values from 73 OASs with a pre-drilling diameter of 1.2 mm were compared between success and failure groups.Results: The success rates of OASs pre-drilled with φ1.2 and 1.5 mm were 94.5% and 83.0%, respectively (P < 0.05); corresponding rates in the mesial zone were 100.0% and 77.3% (P < 0.005), and those in the distal zone were 89.2% and 88.6%, respectively. Placement torques of OASs predrilled with φ1.2 mm in the success and failure groups were 25.9 and 19.2 N·cm, respectively (P < 0.05).Conclusion: A smaller pre-drilling diameter was associated with a higher success rate of OASs in the mid-palatal area, especially in the mesial zone. When pre-drilling diameter of 1.2 mm was used for φ2.0 mm OAS, greater placement torque was indicative of greater OAS stability.
著者
Atsuko Hotta Yasuki Uchida Yasuhiro Namura Mizuki Inaba Mitsuru Motoyoshi
出版者
Nihon University School of Dentistry
雑誌
Journal of Oral Science (ISSN:13434934)
巻号頁・発行日
vol.62, no.3, pp.318-321, 2020 (Released:2020-06-23)
参考文献数
30

This study used finite element (FE) analysis to investigate the stability of miniscrews (screws) placed at the median palate. FE models with variable suture maturity and screw-suture distances were used to examine the relationship with screw stability. Four groups were classified by extent of maturation of the midpalatal suture (0%, 60%, 75%, and 100%). The placement position was set at the center of the suture (0.0 mm), or 0.5, 1.0, and 1.5 mm to the side of the suture, and von Mises stress values in bone and screw displacement were compared among models. The stress value for the unsutured model, in which the screw was placed at the center of the suture, was greater than 30 MPa. Stress values for models in which screws were placed to the side (0.5-1.5 mm) were less than 28 MPa. Maximum screw displacement was greater in the 0.0-mm incomplete suture model than at other placement positions. Because bone conditions vary among patients, placement position and suture maturation should be examined on cone beam-computed tomography images, to ensure screw stability.
著者
Yasuhiro Namura Toshiki Takamizawa Yasuki Uchida Mizuki Inaba Daichi Noma Tomoko Takemoto Masashi Miyazaki Mitsuru Motoyoshi
出版者
Nihon University School of Dentistry
雑誌
Journal of Oral Science (ISSN:13434934)
巻号頁・発行日
vol.62, no.1, pp.48-51, 2020 (Released:2020-01-29)
参考文献数
20
被引用文献数
8

Although there have been improvements in bracket systems precoated with adhesive, removal of adhesive remnants continues to be problematic. This study compared the hardness and maintainability of precoated adhesive with other commercial adhesives. Knoop hardness values were measured after light- or chemical-induced initial curing, immersion in distilled water at 37°C for 24 h and 1,000 and 10,000 thermal cycles after 24 h. Additionally, the forces required to move brackets by 0.5 mm were measured during bracket positioning, and brackets bonded to bovine enamel were examined by field-emission scanning electron microscopy. The Knoop hardness values of the precoated adhesives were lower than those of commercial resin composite adhesives, and hardness was dependent on the amount of filler in the resin matrix. The ability to maintain the device position may depend on the resin matrix composition. Precoated adhesives with less filler and more matrix material are light curable, and remnant resin may be easily removed.
著者
Michiko Ichinohe Mitsuru Motoyoshi Mizuki Inaba Yasuki Uchida Mari Kaneko Reiko Matsuike Noriyoshi Shimizu
出版者
Nihon University School of Dentistry
雑誌
Journal of Oral Science (ISSN:13434934)
巻号頁・発行日
pp.17-0377, (Released:2018-10-26)
参考文献数
35
被引用文献数
19

This study investigated the stability of mini-screws placed in the median palate. The study included 25 patients (7 males, 18 females; mean age, 23.4 ± 5.6 years; age range, 15.0-34.5 years) who had mini-screws placed during orthodontic treatment at Nihon University School of Dentistry Dental Hospital. Mini-screws (diameter, 2.0 mm; length, 9.0 mm) were placed in the median palatal region; the first screw was inserted mesiodistally at the distal contact of the maxillary first molar, and the second screw was placed 6-9 mm mesial to the first screw. Immediately after placement, the placement sites were carefully examined with cone-beam computed tomography and a Periotest device. Screw stability was not related to perforation of the nasal cavity, patient age, or patient sex. The success rate was significantly higher in patients with screw-suture distances of 1.5-2.7 mm than in those with distances of 0-1.4 mm. Moreover, mini-screws could be stabilized when palatal cortical bone thickness was ≥1.5 mm. The success rate was significantly higher in the group with insertion depths of ≥4.5 mm. These results indicate that primary stability of mini-screws requires sufficient cortical bone thickness, insertion depth, and screw-suture distance.
著者
Tomoyasu Kiyokawa Mitsuru Motoyoshi Mizuki Inaba Remi Sano Akari Saiki Go Torigoe Masatake Asano Noriyoshi Shimizu
出版者
日本大学歯学部
雑誌
Journal of Oral Science (ISSN:13434934)
巻号頁・発行日
vol.59, no.3, pp.447-451, 2017 (Released:2017-09-14)
参考文献数
22
被引用文献数
2

The purpose of this experiment was to investigate whether low-intensity pulsed ultrasound (LIPUS) irradiation can inhibit dentoalveolar ankylosis in transplanted rat teeth. LIPUS irradiation (the pulsed ultrasound signal had a frequency of 3.0 MHz, a spatial average intensity of 30 mW/cm2, and a pulse ratio of 1:4) was performed on the face over the re-planted teeth of rats for 4 weeks. After the rats were euthanized, we measured mobility (Periotest value [PTV]) of the transplanted and control teeth using a Periotest. Finally, we performed histological evaluation to detect ankylosis. PTVs tended to be significantly lower for re-planted teeth than for control teeth. Histological evaluation revealed that the roots of all re-planted teeth were coalescent with alveolar bone. Furthermore, no ankylosis was observed in three-fifths of the re-planted teeth following LIPUS irradiation. These results indicate the potential efficacy of LIPUS to inhibit dentoalveolar ankylosis.