著者
加藤 守 千田 浩一 盛武 敬 小口 靖弘 加賀 勇治 坂本 肇 塚本 篤子 川内 覚 松本 一真 松村 光章 大阪 肇 土佐 鉃雄
出版者
公益社団法人 日本放射線技術学会
雑誌
日本放射線技術学会雑誌 (ISSN:03694305)
巻号頁・発行日
vol.70, no.8, pp.814-820, 2014
被引用文献数
1

In recent years, dose justification and optimization have been attempted in percutaneous coronary intervention (PCI); however, deterministic effects have been reported. To prevent radiation skin injuries in PCI, it is necessary to measure the patient entrance skin dose (ESD), but an accurate dose measurement method has not yet been established. In this study, we developed a dosimetry gown that can measure the ESD during PCI using multiple radiophotoluminescence dosimeters (RPLDs). The RPLDs were placed into 84 pockets that were sewn into a dosimetry gown. Patients wear the original dosimetry gown during the procedures, after which we obtain accurate ESD measurements. We believe that this method using RPLDs and a newly-designed dosimetry gown provides accurate ESD measurements during PCI. We expect this system to become a standard method for measuring ESD during PCI.

言及状況

外部データベース (DOI)

Twitter (1 users, 1 posts, 0 favorites)

こんな論文どうですか? 心血管撮影領域における患者被ばく線量測定用装具の開発(加藤 守ほか),2014 https://t.co/nF9Nppl3zP

収集済み URL リスト