- 著者
-
田頭 豊
志村 浩孝
坂本 博
佐藤 和宏
梁川 功
千田 浩一
- 出版者
- 医用画像情報学会
- 雑誌
- 医用画像情報学会雑誌 (ISSN:09101543)
- 巻号頁・発行日
- vol.33, no.2, pp.43-47, 2016-06-27 (Released:2016-07-01)
- 参考文献数
- 10
June 7, 2015, Japan Network for Research and Information on Medical Exposures(J-RIME)published Diagnostic Reference Levels(DRLs 2015)for the first time in Japan. To operate the DRL, it is necessary to collect a great deal of radiation dose information of own facilities efficiently. The purpose of this study was to construct a system for collecting dose information in CT examination using Radiation Dose Structured Report(RDSR). As a result of displaying the contents of RDSR using the free software, it was confirmed that there are differences in the information in RDSR by equipment and manufacturers. In-house software of dose management could be designed by focusing on the enumerated value, it was possible to extract information of interest from RDSR. For facilities which cannot introduce a commercial system, development of software for collecting radiation dose information using RDSR contributes greatly to the operation of DRL.