著者
小宮 勲 白坂 崇 梅津 芳幸 橘 昌幸 泉 隆
出版者
公益社団法人 日本放射線技術学会
雑誌
日本放射線技術学会雑誌 (ISSN:03694305)
巻号頁・発行日
vol.60, no.2, pp.270-277, 2004
参考文献数
20
被引用文献数
12 14

In this study, we investigated the usefulness of the fluorescent glass dosimeter for measuring patient dose. The fluorescent glass dosimeter is constructed of a glass element and its holder. One type has a tin (Sn) filter and the other does not. The characteristics of these two types of fluorescent glass dosimeters were studied in the range of diagnostic X-ray energy. The result was excellent for each characteristic. Directional dependency, however, was recognized in the fluorescent glass dosimeter with tin (Sn) filter. Based on these evaluations, patient skin dose was measured for abdominal interventional radiology and diagnostic digital subtraction angiography using the holder without filter, which is less direction-dependent and eliminates obstructive shadows in radiography and fluoroscopy. The average skin dose of 30 patients for abdominal IVR was 1.17±0.44 Gy (0.51-1.94 Gy), while those for diagnostic DSA examination was 0.54±0.21 Gy (0.15-1.02 Gy). The fluorescent glass dosimeter provides high capability for skin dose measurement. The fluorescent glass dosimeter is also useful for controlling patient dose during IVR procedures.
著者
坂本弘巳 長 哲二 斎藤 高志 小宮 勲 泉 隆
出版者
九州大学
雑誌
九州大学医療技術短期大学部紀要 (ISSN:02862484)
巻号頁・発行日
vol.23, pp.47-50, 1996-03
被引用文献数
2

When linear accelerator exceeds 10MV, it produces photonuclear reaction, and generates continuous energy. Measuring devices are very few in number and determination of energy spectrum is very difficult. This paper is designed to show the result of survey made by bubble detectors.