In this study, we investigated the usefulness of the fluorescent glass dosimeter for measuring patient dose. The fluorescent glass dosimeter is constructed of a glass element and its holder. One type has a tin (Sn) filter and the other does not. The characteristics of these two types of fluorescent glass dosimeters were studied in the range of diagnostic X-ray energy. The result was excellent for each characteristic. Directional dependency, however, was recognized in the fluorescent glass dosimeter with tin (Sn) filter. Based on these evaluations, patient skin dose was measured for abdominal interventional radiology and diagnostic digital subtraction angiography using the holder without filter, which is less direction-dependent and eliminates obstructive shadows in radiography and fluoroscopy. The average skin dose of 30 patients for abdominal IVR was 1.17±0.44 Gy (0.51-1.94 Gy), while those for diagnostic DSA examination was 0.54±0.21 Gy (0.15-1.02 Gy). The fluorescent glass dosimeter provides high capability for skin dose measurement. The fluorescent glass dosimeter is also useful for controlling patient dose during IVR procedures.
著者
粟井一夫 浅田 朋宏 天内 廣 梅津 芳幸 水谷 宏 梁川 功 山下 一也 伊知地 宏志
出版者
公益社団法人日本放射線技術学会
雑誌
日放技学誌 (ISSN:03694305)
巻号頁・発行日
vol.48, no.1, pp.83-106, 1992
被引用文献数
3

これは過去1年間におけるX線映画領域でのセンシトメトリー班の活動報告である.第47回総会において, その活動報告および公開討論会を行った.そこでの会員からの意見と助言は, この報告書を作成するにあたって参考にさせていただいた.施設間で同じ結果を得ることは, 現在の現場の状況を考慮するとなかなか難しいものがある.そのような状況で少しでも正確な結果を得るためにはどうすべきかを第一番に考えてこのような標準化をはかった.正確な特性曲線作成には現像処理系および濃度測定が大きな影響を及ぼしている.センシトメトリーの標準化をはかるにはこの二点の規格化が不可欠である.濃度計に関しては, JIS・IEC委員会「医用X線写真濃度計班」の報告に期待したい.終りに班活動を遂行するにあたり, 終始援助と理解をいただいた各施設の技師長と技師諸兄, そして実験材料を無償で提供していただいた日本コダック(株)および富士メディカル(株)に深謝を表する次第である.また, データ整理にあたり特性曲線プログラムを無償で提供していただいた斉藤誠氏(東京都立池袋保健所)に感謝いたします.