著者
松田 忠義 吉田 翼 有岡 孝則 高塚 貞雄 二川 佳央 菊地 眞
出版者
Japanese Society for Thermal Medicine
雑誌
日本ハイパーサーミア学会誌 (ISSN:09112529)
巻号頁・発行日
vol.4, no.4, pp.317-329, 1988-12-01 (Released:2009-10-21)
参考文献数
21
被引用文献数
5 5

Heating experiments on phantoms and miniature pigs were made by using two lens applicators and a waveguide applicator fed by the 430MHz microwave heating system (HTS-100, producted by Tokyo Keiki Co., Ltd.). Temperature distributions of agar phantoms and miniature pigs were measured at regular intervals during heating. As the results of these experiments, the followings are obtained.1. The results of effective heating depth varying with a focal length confirm lens effect.2. Using a four-aperture lens applicator and miniature pigs, the heating depth of 6cm and it almost doubles that of the ordinary waveguide applicators. The heating area is 12cm x 6cm at 4cm depth.3. Dielectric properties of miniature pig buttocks are nearly equal to these of the agar phantom containing 0.35% NaCl and 0.05% NaN3.
著者
副島 健市 小柳 仁 吉田 翼 有岡 孝則 大内 田明
出版者
JAPANESE SOCIETY FOR ARTIFICIAL ORGANS
雑誌
人工臓器 (ISSN:03000818)
巻号頁・発行日
vol.14, no.4, pp.1782-1784, 1985

開心術を行なう際, 人工心肺回路にopen circuitを用いるかぎり, つねに空気塞栓の危険を伴ってくる. 空気塞栓予防対策は, 以前より種々の方法があるが, それぞれ一長一短があった. 今回われわれは超音波を用い, 貯血槽の液面を感知するレベル・センサーを試作したので, 使用経験を報告する. 装置はWilliam Harvey (WH) 1300, 1500用として, 液面に対し直下より垂直方向に用いるものと, WH 1700用として, 液面に平行して横から用いる, 2種類を試作した. いずれも液面の低下を警報で知らせ, さらに危険域まで低下すれば, 警報と同時に, 送血ポンプの非常停止を行なう. 本装置はノイズ対策が十分で, また光や静電気などで誤作動する心配がなく, 確実に作動するため, きわめて有用であった.