pp.tn.2021-0130, (Released:2022-03-23)
参考文献数
13

We have developed a new device, consisting of a 3-cm RF coil and an immobilizer, to acquire high-resolution MR images of the testis. With the approval of our institutional review board, we conducted an MRI study on a cohort of healthy volunteers to test this device. With the participants in the supine position, we placed the dedicated immobilizer and RF coil on the scrotum for typically no more than 3 min. Subsequently, T2-weighted images were acquired with an in-plane resolution of 117 µm using a 3-T MR scanner and the periodically rotated overlapping parallel lines with enhanced reconstruction (PROPELLER) sequence. The total scan time ranged from 12 to 30 min (average 20 min). High-resolution MR images of the testis were acquired without deterioration by motion artifacts. Our results showed that the combined use of a small RF coil and an immobilizer is a feasible option for acquiring high-resolution MR images of the testis.