著者
三宅 仁 藤正 巌 井街 宏 西坂 剛 大道 久 岩井 矩成 河野 明正 渥美 和彦
出版者
一般社団法人 日本人工臓器学会
雑誌
人工臓器 (ISSN:03000818)
巻号頁・発行日
vol.7, no.1, pp.145-148, 1978-01-15 (Released:2011-10-07)
参考文献数
4

According to our total artificial heart experiment, the long survival can be achieved by maintainig of cardiac output in normal rate, which is 80 to 100ml/kg.When an experimental animal is made exercise, its cardiac output should be increase. The amount of increased cardiac output is seemed to relate venous pressure, peripheral resistance or partial O2 pressure of blood. The relation between load of exercise and such physiological data, as CVP, TPR and O2 consumption, were analysed quantitatively in this report.The results were; 1) By CVP, the cardiac output of artificial goat during exercise could not be determined. 2) TPR has some relation to the cardiac output but we need more precise study. 3) O2 consumtion is seem to be most useful information to control total artificial heart.
著者
阿部 裕輔 磯山 隆 小野 稔 斎藤 逸郎 三浦 英和 鎮西 恒雄 望月 修一 磯山 隆 小野 稔 鎮西 恒雄 斎藤 逸郎 三浦 英和 川崎 和男 望月 修一 井街 宏 バスク ジャロミル ドブザク ペーター 中川 英元 光宗 倫彦 河野 明正 小野 俊哉 時 偉 杉野 礼佳 井上 雄介 岸 亜由美
出版者
東京大学
雑誌
基盤研究(A)
巻号頁・発行日
2005

無拍動流完全人工心臓の可能性を検討するために、体内埋込式無拍動流完全人工心臓を開発し、成獣ヤギに埋め込んで研究を行った結果、無拍動流でも生理的制御(正常な状態に維持するための制御)が可能で、ヤギの一般状態、血行動態、肝機能、腎機能や自律神経機能を正常に保つことができることを明らかにした。しかし、無拍動流では人工心臓への十分な血液流入を維持できないことがあったため、体内埋込式完全人工心臓に生理的制御を適応する場合には、ある程度の拍動性が必要であると考えられた。
著者
阿部 裕輔 鎮西 恒雄 磯山 隆 満渕 邦彦 松浦 弘幸 馬場 一憲 河野 明正 小野 俊哉 望月 修一 孫 艶萍 今西 薫 吉澤 誠 田中 明 内山 賢一 藤正 巌 渥美 和彦 井街 宏
出版者
JAPANESE SOCIETY FOR ARTIFICIAL ORGANS
雑誌
人工臓器 (ISSN:03000818)
巻号頁・発行日
vol.26, no.1, pp.21-26, 1997-02-15
被引用文献数
7

1/R制御により、完全置換型人工心臓ヤギで532日の長期生存を得た. 術後経過としては、ポンプ内に血栓を生じたため、左右の血液ポンプを交換したこと、胸部の圧迫壊死層から出血をきたし、貧血の状態となったことなどがあったが、制御は順調に継続できた. 生存期間を通して、血行動態は安定し、右心房圧も低値に保たれていた. 血液生化学データ上も異常は見られず、ホルモン値にも異常は見られなかった. 剖検時に腹水はなく、肉眼的には肝臓病変も見られなかった. また、心拍出量は、大動脈圧とは関係が見られず、総コンダクタンスと相関が見られ、術後2週間および出血後2週間はヘマトクリット値と逆比例の関係が見られたが、それ以外の時期にはヘマトクリット値や総タンパク値とはあまり関係がないなどの興味深い所見も得られた. 本研究により、1/R制御の生理的な長期安定性が確認できた.