著者
三宅 仁 藤正 巌 井街 宏 西坂 剛 大道 久 岩井 矩成 河野 明正 渥美 和彦
出版者
一般社団法人 日本人工臓器学会
雑誌
人工臓器 (ISSN:03000818)
巻号頁・発行日
vol.7, no.1, pp.145-148, 1978-01-15 (Released:2011-10-07)
参考文献数
4

According to our total artificial heart experiment, the long survival can be achieved by maintainig of cardiac output in normal rate, which is 80 to 100ml/kg.When an experimental animal is made exercise, its cardiac output should be increase. The amount of increased cardiac output is seemed to relate venous pressure, peripheral resistance or partial O2 pressure of blood. The relation between load of exercise and such physiological data, as CVP, TPR and O2 consumption, were analysed quantitatively in this report.The results were; 1) By CVP, the cardiac output of artificial goat during exercise could not be determined. 2) TPR has some relation to the cardiac output but we need more precise study. 3) O2 consumtion is seem to be most useful information to control total artificial heart.
著者
阿部 裕輔 鎮西 恒雄 磯山 隆 満渕 邦彦 松浦 弘幸 馬場 一憲 河野 明正 小野 俊哉 望月 修一 孫 艶萍 今西 薫 吉澤 誠 田中 明 内山 賢一 藤正 巌 渥美 和彦 井街 宏
出版者
JAPANESE SOCIETY FOR ARTIFICIAL ORGANS
雑誌
人工臓器 (ISSN:03000818)
巻号頁・発行日
vol.26, no.1, pp.21-26, 1997-02-15
被引用文献数
7

1/R制御により、完全置換型人工心臓ヤギで532日の長期生存を得た. 術後経過としては、ポンプ内に血栓を生じたため、左右の血液ポンプを交換したこと、胸部の圧迫壊死層から出血をきたし、貧血の状態となったことなどがあったが、制御は順調に継続できた. 生存期間を通して、血行動態は安定し、右心房圧も低値に保たれていた. 血液生化学データ上も異常は見られず、ホルモン値にも異常は見られなかった. 剖検時に腹水はなく、肉眼的には肝臓病変も見られなかった. また、心拍出量は、大動脈圧とは関係が見られず、総コンダクタンスと相関が見られ、術後2週間および出血後2週間はヘマトクリット値と逆比例の関係が見られたが、それ以外の時期にはヘマトクリット値や総タンパク値とはあまり関係がないなどの興味深い所見も得られた. 本研究により、1/R制御の生理的な長期安定性が確認できた.
著者
井街 宏 藤正 巌 大道 久 真野 勇 西坂 剛 岩井 矩成 渥美 和彦
出版者
一般社団法人 日本人工臓器学会
雑誌
人工臓器 (ISSN:03000818)
巻号頁・発行日
vol.5, no.6, pp.321-327, 1976-12-15 (Released:2011-10-07)
参考文献数
16

The rapid decrease in hematocrit (Ht) and total protein (Tp), gradual increase in central venous pressure, and peripheral circulatory insufficiency such as impossibility of standing up, poor appetite, anuria or oliguria and constipation or mucous feces were the most serious physiological problems in recent total artificial heart (AH) animal experiments. For these phenomena, a phsiological hypothesis was presented by authors under the consideration of that too much cardiac output for animal might induce such phenomena.As the results of animal experiments with goat, the following finidings were obtained.(1) High cardiac output of AH causes a marked decrease in Ht and Tp, as well as peripheral circulatory insufficiency.(2) By maintaining the cardiac output within the range of 80 to 100ml/Kg/min. during the experiments, maximum and average survival days of the goats replaced totally with AH were markedly prolonged, from 10 days to 54 days and from 4.5 days to 30.5 days, respectively. And the surviving condition of these goats were almost normal physiologically and pathophysiologically.(3) As conclusion, it became obvious that the cardiac output was one of the most important physiological parameter for control of the circulatory system.
著者
宮本 晃 長谷川 隆光 北村 信三 梅田 正五 川野 幸志 進藤 正二 陸川 秀智 塩野 元美 小笠原 弘二 並木 義夫 折目 由紀彦 瀬在 幸安 渥美 和彦 藤正 巌 井街 宏
出版者
JAPANESE SOCIETY FOR ARTIFICIAL ORGANS
雑誌
人工臓器 (ISSN:03000818)
巻号頁・発行日
vol.15, no.2, pp.604-607, 1986

開心術後のLOS症例と, 体外循環からの離脱困難症例2例に補助循環として補助人工心臓を使用した。症例1は61才女性, AR+MRの診断にて二弁置換術を施行, 症例2は23才女性, MR+TRの診断にて僧帽弁置換術と三尖弁弁輪形成術を施行, 症例3は急性心筋梗塞症に心室中隔穿孔と左室瘤を合併し, 穿孔部閉鎖術と瘤切除を施行した。このうち2例は離脱可能であったが, 3例とも救命できていない。補助人工心臓の効果についての異論はないが, 成績向上のためには補助心臓シヌテム運用面における問題点を解決する必要があり, 本シヌテムの改良点を含めて経験症例を検討した。補助心臓ポンプの容量およびカニューレのサイズは, 抗血液凝固剤の使用を減少させる意味で最小限とした。適応基準については既に確立されているが, 運用面では補助心臓適用時期の問題, 補助心臓とIABP併用の際の駆動時相の問題, さらに補助循環中の心拍出量等について解析を加えた。